Dräger Polaris 600 Instructions For Use Manual page 71

Surgical light
Hide thumbs Also See for Polaris 600:
Table of Contents

Advertisement

LED
Status LED (D)
Polaris 600 remote control for MedView
camera
Fault
Cause
Remote control is not within the reception range of the IR
receiver
Remote control is
not responding
(status LED lights
up green when
key is pressed)
Room address of the remote control or of the MedView
camera is incorrectly set
Battery is discharged
Remote control is
not functioning
(status LED does
Battery is defective
not light up when
key is pressed)
Instructions for use Polaris 600
Status
Off
Continuously white
on the light control
panel
Continuously orange
on the wall-mounted
control panel
Cause
Ambient light mode (Endo light)
is not switched on (OR light
mode is switched on)
Light is in standby mode
Light is in ambient light mode
(Endo light)
Troubleshooting
Remedy
No remedial action nec-
essary
No remedial action nec-
essary
Remedy
Aim the remote control at
the IR receiver, see
"Operating the MedView
camera using the
Polaris 600 remote con-
trol for MedView camera"
on page 64
Hold the remote control
closer to the IR receiver,
see "Operating the Med-
View camera using the
Polaris 600 remote con-
trol for MedView camera"
on page 64
Remote control and
MedView camera must
be set to the same room
address.
Notify DrägerService
Fit a new battery, see
"Replacing the battery in
the Polaris 600 remote
control for MedView
camera" on page 65
Fit a new battery, see
"Replacing the battery in
the Polaris 600 remote
control for MedView
camera" on page 65
71

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents