Safety-related information
WARNING
Risk due to electromagnetic disturbance
– Maintain a distance of at least 0.3 m (1.0 ft)
between this device and wireless
communication devices to ensure that the
essential performance of this device is
fulfilled.
– Maintain an adequate distance between
this device and other medical electrical
equipment.
Radio interference
WARNING
Risk of device malfunction
The light system can cause radio-frequency
interference.
Shielding of components or reorientation of
the light system may be necessary.
Installing accessories
CAUTION
Risk of device failure
Install the accessory on the basic device in
accordance with the instructions of the basic
device.
Check for secure connection to the basic device.
Strictly observe the instructions for use and
assembly instructions.
Storing the instructions for use
WARNING
Risk of incorrect use
Instructions for use must be kept accessible
to the user.
10
Training
Training for users is available via the Dräger
organization responsible (see www.draeger.com).
Functional safety
Light system
The essential performance of lights consists in:
– illumination of the operating field,
– reduction of thermal radiation into the operating
field.
Display mounts
The essential performance characteristics of
display mounts are:
– the mounting of up to 2 displays.
Mandatory reporting of incidents
Serious incidents with this product must be
reported to Dräger and the responsible authorities.
Instructions for use Polaris 600