Dräger Polaris 600 Instructions For Use Manual page 72

Surgical light
Hide thumbs Also See for Polaris 600:
Table of Contents

Advertisement

Troubleshooting
Fault
Cause
Remote control is
Membrane keypad is damaged
not functioning
(status LED is lit
green continu-
ously)
Remote control is
Battery charge is exhausted
functioning but
status LED
flashes green
when key is
pressed
Wireless video transmission
Fault
Cause
MedView camera is carrying out an initialization after
standby mode.
Connection between the radio transmitter and the radio
receiver is interrupted
Connection between the radio transmitter and the radio
receiver is interrupted
Video image is
not displayed
Video image is
Unknown
not displayed and
cannot be rees-
tablished
72
Remedy
Replace defective
remote control
Fit a new battery, see
"Replacing the battery in
the Polaris 600 remote
control for MedView
camera" on page 65
Remedy
No remedial action nec-
essary. This procedure
takes up to 30 seconds
Switch the MedView
camera to standby mode.
Switch the MedView
camera on again after a
few seconds
Reestablish the connec-
tion:
1 Push up the locking
plate on the radio
receiver housing (dis-
play mount).
2 Set the on/off switch
on the radio receiver
to 0.
3 Wait for approx. 5
seconds.
4 Set the on/off switch
on the radio receiver
to I.
 The connection will
be reestablished.
This procedure takes
up to 30 seconds
Notify DrägerService
Instructions for use Polaris 600