Output signals

NOTE: Only analogue sine and cosine outputs are functionally safe.
Readhead output
Function
Output type
Power
-
Cosine
Incremental signals
Analogue
Sine
Reference mark
Analogue
Limits
Open collector
Set-up
-
Calibrate
-
-
Shield
-
1
There is no inner shield on UHV cables.
Original instructions
Signal
Colour
5 V Power
Brown
0 V Power
White
+
Red
V
1
Blue
+
Yellow
V
2
Green
+
Violet
V
0
Grey
V
Pink
p
V
Black
q
V
Clear
X
CAL
Orange
1
Inner shield
Green / Yellow
Outer shield
Outer screen
Ti0000 interface output
Function
Output type
Power
-
Cosine
Incremental signals
Analogue
Sine
Reference mark
Analogue
Limits
Open collector
Set-up
-
Calibrate
-
-
Shield
-
Signal
Pin
5 V Power
4
5 V Sense
5
0 V Power
12
0 V Sense
13
+
9
V
1
1
+
10
V
2
2
+
3
V
0
11
V
7
p
V
8
q
V
6
X
CAL
14
Inner shield
Not connected
Outer shield
Case
Ti0000 interface
output connector: