Reanalysis Layout
2028451-182B
b. When the correct point is selected, use the trimpad to adjust its
position.
c.
To verify correct positioning, refer to the values in the
Measurement Legend in the lower left corner of the display.
For more information on the Measurement Legend, refer to
"Reanalysis Layout"
d. Repeat step a through step c for each fiducial point to be
adjusted.
5. When you are done adjusting the fiducial points, do one of the
following:
To discard your adjustments and start over, press the F4 key.
The original readings are restored. Return to step 2 to start over.
To save your adjustments, press the F5 key.
The changes are saved
6. Repeat from step 2 to make adjustments in the other edit mode.
7. After all your changes have been made, press the F6 key to return to
the original menu options.
Selecting the Reanalysis option after acquiring a resting ECG displays
the following screen. The screen's key features are described in the
following table.
MAC™ 1600
on page 5-9.
Recording a Resting ECG
047A
5-9