GE MAC 1600 Operator's Manual page 100

Ecg analysis system
Hide thumbs Also See for MAC 1600:
Table of Contents

Advertisement

System Configuration
9-8
Field
12 leads: 4x3
Display Format
Display Lead Group
Page 2
Printer Leads
Start rhythm report on new
page
Pace Enhancement
Hookup Advisor
MAC™ 1600
Comment
Enables/disables a display option that shows four three-
waveform columns. Available only if the R12L system
option is enabled.
Selects the display format of the resting ECG. Default
value is 3 leads: 1x3. Other values depend on which of the
previous four fields are set.
Determines which group of leads is displayed. The
available values depends on which Display Format is
selected. For example, if 3 Leads: 1x3 is selected, the
available values are:
3 rhythm leads
1st group
2nd group
3rd group
4th group
If either of the 12-lead display formats is selected, this field
is not available, since all 12 leads are displayed.
Identifies the default set of leads used for printing. The
values are:
First 6
Second 6
Rhythm 6
12
Determines whether the rhythm report should begin on a
its own page.
Increases the readability of pacemaker ECG either by
augmenting small pace pulses or by truncating large pace
pulses. If enabled, pace enhance is done in two steps: (1)
Add a marker (1.5mV amplitude, 6 ms duration) to the
electrode signal. (2) Limit the sum to 0.5mV in the lead
signal.
Enables/disables the Hookup Advisor option, which
visually indicates the quality of lead signals. For more
information, refer to
"Hookup Advisor Module"
page 5-12.
on
2028451-182B