Maintenance
Cleaning Cables and Leadwires
Disinfecting Cables and Leadwires
10-6
Never autoclave or steam clean cables or leadwires.
Always wipe gently to avoid pulling long wires from the connectors.
Always remove cables and leadwires from the device before cleaning.
Failure to observe these cautions could result in damage to the contact
metal ends, thereby affecting signal quality.
Use the following procedure to clean the cables and leadwires.
NOTE
Cleaning removes dirt and marks but does not disinfect.
1. Dilute mild dishwashing detergent in water to create a cleaning
solution.
2. Soak a clean cloth in the solution and wring out any excess.
3. Thoroughly wipe the exterior of the cables and leadwires with the
damp cloth.
4. Repeat step 2 and step 3 as necessary until adequately cleaned.
5. Wipe with a dry, clean cloth or paper towel and let air dry.
Use the following procedure to disinfect the cables and leadwires.
NOTE
Clean and dry the cables and leadwires before disinfecting them.
1. Use a lint-free cloth or wipe with a solution of Sodium Hypochloride.
2. Wring excess liquid from the cloth.
3. Gently wipe the cabling.
4. Wipe off the disinfectant with a clean, lightly moistened cloth.
NOTE
If fluid pools around the connectors, blot dry with a soft, lint-free
cloth.
5. Wipe with a dry, lint-free cloth and let air dry for at least 30 minutes.
NOTE
Drying times vary based on the environmental conditions.
DO NOT use excessive drying techniques, such as ovens, forced
heat, or sun drying.
MAC™ 1600
2028451-182B