7.

Sensitivity Adjustment / Calibration Test

7.1. Automatic sensitivity adjustment

The sensitivity of the balance is automatically adjusted using the internal mass according to ambient
temperature change, set time, or interval time. For details, refer to the separate instruction manual listed
in "1.1. Detailed Manuals".

7.2. Sensitivity adjustment using the internal mass

The sensitivity adjustment / calibration test screen is displayed with the
screen.
2. Internal
sensitivity
adjustment
3. Internal
calibration
test
Name
External sensitivity
1
adjustment
Internal sensitivity
2
adjustment
3
Internal calibration test
4
External calibration test
Do not allow vibration or drafts to affect the balance during sensitivity adjustment / calibration test.
For details, refer to the separate instruction manual listed in "1.1. Detailed Manuals".
Your external calibration weight is used to adjust the balance.
Instructions are displayed on the screen for how to adjust the
sensitivity.
The internal mass is used to automatically adjust the balance.
The internal mass is used to check the accuracy of weighing, and
the result is output.
Note that the balance is not adjusted.
Your external calibration weight is used to check the accuracy of
weighing, and the result is output.
Note that the balance is not adjusted.
14
button on the weighing
Description
1. External
sensitivity
adjustment
4. External
calibration
test