1500°C Tube Furnace

Tube Adapters

CAUTION : Do not operate the furnace without properly sized and installed tube
adapters.
Install tube adapters to each end of the furnace, through the tube adapter at the end,
through the furnace chamber below careful. Do not touch / Burn the elements, allow and
push through tube adapter on opposite end of furnace.
Similar to the element in Section "Heating Elements", pack ceramic wool into any gaps
between the outside of the process tube and the tube adapter.
NOTE Only an authorized or a qualified person shall place the process tube in the device while
in operation. Do not place your hand inside the unit while placing the process tubes. If
misused, the person may come in contact with live electrical parts and touching the live parts
can result in an electrical shock, burn, or electrocution. It is always recommended to have
appropriate electrical PPE while operating the device; insulated TOOLS; insulating clothing;
OPERATOR standing on an insulating surface etc.