Over Temperature Controller Operation - Thermo Scientific STF55433 Installation And Operating Manual

Lindberg/blue m 1500c general-purpose tube furnaces
Table of Contents

Advertisement

| Operation - 3216i Over Temperature Controller
Chapter 8
|
8-72
305435H03
Table 5 Default settings for the Excess temperature controller
(STF55433BC-1, STF55433BCOMC-1)
Parameter
Description
ALARM.1.Threshold
Excess temp threshold 1550°C
ALARM.1.Hysteresis
Alarm Hysteresis
INPUT.Units
Display Units
INPUT.PVInValue
PV Input Value
ACCESS.HomeDisplay
Home Display
INPUT.PeakHigh
Peak High
INPUT.PeakReset
Peak Reset
INPUT.TimeUnits
Time Units
INPUT.Time
Time exceeded
INPUT.PVOffset
PV Offset
Customer ID
Customer ID
NOTE *Level 1+2 Read Only states that, Level 1 gives Read only access to user
where as Level 2 gives Write access along with Read access.
NOTE To enter Level 2 and Level 1 refer to the section "To Enter Level 2" .

Over Temperature controller Operation

When the controller is turned ON it will perform a short self-test and then display a default
page as shown in the below image. The excess temperature Alarm threshold (setpoint) is
found in the display. This excess temperature controller will be configured with respect to
its functionality in the factory.
Level
Access
Level 1+2
Read/Write
1°C
Level 2 Read only
°C
Level 1+2
Read/Write
Displays Process
Level 1+2 Read only
Value
Excess temp
Level 1+2
threshold (1550°C)
Read/Write
Peak High PV value
Read only
Peak Reset
Level 1+2
Read/Write
Mins
Level 1+2
Read/Write
Time in alarm
Level 1+2
Read/Write
+/- adjusted with
Level 1+2
respect to main
Read/Write
controller PV read-
ing
1502
Level 2 Read only
1500°C Tube Furnace