| Main Controller: 25x500 Segment Programmable
Chapter 7
|
7-62
305435H03
5. Change the Tune value ON with UP button.
6. Press PAGE + SCROLL buttons to start the Auto tune process.
7. Controller begin the Auto tuning process.
8. Once after Auto tune, new PID value will be loaded in to the controller with respective
to the tuned SP value.
NOTE In the Factory, PID values are Auto tuned to 550°C.
It is recommended to tune to desire Setpoint for best accuracy.
After Auto tune, PID value will be changes with respect to tuned setpoint value.
1500°C Tube Furnace