1500°C Tube Furnace
10. Press SCROLL to select Segment1 Target SP, modify Target SP as per the
requirement with the Lower or Raise button, the controller will accept the new value
which is indicated by a brief flash of the display.
11. Press SCROLL to select Ramp Rate, its default value is 0.1 (Ramp rate/Sec). It can be
altered as per the requirements with the Lower or Raise button, the controller will
accept the new value which is indicated by a brief flash of the display.
12. Press SCROLL to select Holdback Type, its default value is Off.
13. Press SCROLL to select Event Outs, Skip all Event outs by pressing SCROLL button.
14. Continue the program by entering next Segment values as shown in the above steps.
Main Controller: 25x500 Segment Programmable |
Chapter 7
305435H03 | 7-59