Thermo Scientific STF55433 Installation And Operating Manual page 69

Lindberg/blue m 1500c general-purpose tube furnaces
Table of Contents

Advertisement

1500°C Tube Furnace
4. Press SCROLL button to select Alarm2 Low threshold value.
5. Press RAISE/LOWER buttons to change Alarm2 Low Threshold value, the controller
will accept the new value and is indicated by a brief flash of the display.
6. Press SCROLL button to select Alarm2 Hysteresis value.
7. Press RAISE button to change Alarm2 Hysteresis value, the controller will accept the
new value and is indicated by a brief flash of the display.
NOTE Alarm1 (Absolute high) is read only; Alarm2 and Alarm3 are L2 access alarms.
Main Controller: 25x500 Segment Programmable |
Chapter 7
305435H03 | 7-67