nCounter Pro Analysis System User Manual
NOTE:
The Prep Station checks electrode function during the validation steps of the process
before any samples are processed. If it detects problems with an electrode, the processing will
halt and wait for user intervention. The screens described below will then appear, allowing the
user to fix the bent electrodes.
1.
Select align electrodes on the System
Hardware Menu.
The 'Align Electrodes—STEP 1' screen will
appear (Figure 114).
2.
If the electrode is severely bent such that a
cartridge cannot be inserted, first close the
electrode fixture without a cartridge in it and
adjust the position of the electrode such that it
is positioned similarly to the other electrodes.
If the electrode is only slightly bent and a
cartridge
maneuvering the bent electrode, then insert a
cartridge before closing the electrode fixture.
Press next.
The 'Align Electrodes Menu—STEP 2' screen will
appear (Figure 115).
FOR RESEARCH USE ONLY. Not for use in diagnostic procedures.
76
can
be
inserted
by
Figure 114: The 'Align Electrodes—STEP 1' screen.
carefully
Figure 115: The 'Align Electrodes—STEP 2' screen.
SEV-00184-01