nCounter Pro Analysis System User Manual
13.
The 'Hybridized Samples' screen will appear (Figure 29). Lift the metal lid to place the hybridized
sample strip tube in the sample strip holder, ensuring that tube 1 aligns with position 1
(Figure 30). Note that the strip tube is asymmetrically keyed, and if the strip tube is placed
incorrectly, the lid won't close properly, and the Prep Station will not be able to start processing.
Close the metal lid and press next.
Hybridized sample strip tubes have two notches to ensure proper orientation (between tubes 1-
2 and 8-9).
IMPORTANT:
placing tubes on the deck. Leaving the caps on will result in a pause in the protocol that
requires user intervention.
Figure 29: The 'Hybridized Samples' screen
FOR RESEARCH USE ONLY. Not for use in diagnostic procedures.
24
Ensure that all tube caps are removed from the hybridized samples prior to
.
Figure 30: Hybridization sample strip tubes with
notched orientation guides.
SEV-00184-01