nanoString nCounter Pro User Manual page 28

Analysis system
Table of Contents

Advertisement

nCounter Pro Analysis System User Manual
1.
Press next. The 'Run Successfully Completed'
screen will appear (Figure 37).
2.
The 'Run Successfully Completed' screen lists the
steps to follow once the run is complete,
including:
a.
Remove and discard empty reagent plates.
b.
Remove and discard the empty tip racks and
foil piercers.
c.
Remove and discard the sample strip.
d.
Remove the sample cartridge and seal the
wells.
3.
To release the fixture after the run is complete, lift the lever in the center top of the device
towards the front with a finger as shown in Figure 38.
4.
After processing is complete, it is important to
do the following:
a.
Seal the wells immediately with the adhesive
film provided to prevent evaporation.
b.
Samples should be protected from as much
light as possible. Store them in the dark to
prevent photobleaching.
c.
Store samples at 4°C. Once sealed, samples
can be stored at 4°C for up to a week with
minimal degradation.
d.
Empty the waste containers.
5.
Press finish to return to the Main Menu.
FOR RESEARCH USE ONLY. Not for use in diagnostic procedures.
28
Figure 37: The 'Run Successfully Completed'
screen.
Figure 38: Releasing the fixture after a
completed run.
SEV-00184-01