nanoString nCounter Pro User Manual page 23

Analysis system
Table of Contents

Advertisement

SEV-00184-01
IMPORTANT:
position of the cartridge slightly. Make sure that the electrodes are correctly aligned. If the
electrodes are not aligned, use the 'Align Electrodes' workflow under the Maintenance
menu. The Prep Station will not be able to process any of the samples if there are bent
electrodes.
Figure 26: The 'Electrode Fixture' screen.
12.
The 'Empty Strip Tubes' screen will appear
(Figure 28). Place the empty strip tubes on
the deck.
Press next.
IMPORTANT:
provided by NanoString. Other tubes have
different dimensions and will cause system
failure.
If any resistance is felt while lowering the fixture, stop and adjust the
Only
use
strip
Figure 27: Carefully lower the electrode
fixture into place over the cartridge, until it
clicks.
tubes
Figure 28: The 'Empty Strip Tubes' screen.
FOR RESEARCH USE ONLY. Not for use in diagnostic procedures.
nCounter Pro Analysis System User Manual
23