Troubleshooting/Eliminating Errors
Please note:
For pumps with level switch and early warning:
In this case, air can normally neither enter the pipe system nor the
liquid end as the pump switches to early warning when the first
switching point is reached: when the second switching point is
reached (after a path of 30 mm) the pump switches itself off.
14.1.4 Red warning lamp (8) lights up - "Error" readout on display
(4) flashes
Lack of chemicals or function plug
Remedy:
• The error message acknowledges itself by filling up the supply
container.
• Check whether the function plug level (11) is plugged in.
No or too little metering
Remedy :
• Briefly press the
Stop" (remote on/off).
• If the error message appears again bleed the liquid end and increase
the stroke length.
System error
Remedy:
• Briefly press the
Stop" (remote on/off)
> Also see chapter 15 / page 82 for information about this.
P
key or execute the control function "Pause/
P
key or execute the control function "Pause/
Version 2.0