Repair and Replacement
6.5.3 REPLACING THE MAIN CIRCUIT BOARD
Part #502029
Included with the Replacement Kit:
• Main Circuit Board
• Six-pin Jumper
Additional Supplies Required:
• Phillips Screwdriver
• Slotted Screwdriver (smaller than 1/8")
• Retain 6-pin jumper from old board
WARNING:
Electrical Shock Hazard: Disconnect the electrical supply before making repairs to the unit.
CAUTION:
Electronic components used in this unit are subject to damage by static electricity. Use proper
static discharge and grounding precautions when servicing the equipment.
STEP 1
Remove the cover according to the instructions in Section 5.
STEP 2
Remove the circuit board from the groove in the unit base, and disconnect the four wire connectors
and the 8-pin jumper.
STEP 3
Remove the new circuit board from the static protective bag. Connect the four wire connectors and
the 6-pin jumper to the new circuit board as shown:
6-22
(REMstar Choice LS and Reliance)
Empty
(2 pin)
Jumper
(6 pin)
J4
Power Switch
Connector
(2 Pin)
Motor
Connector
J1
(8 Pin)
Empty
(2 Pin)
J2
DC Inlet
Wires
(2 Pin)
Transformer
Wires
(4 Pin)
REMstar Service Manual