6.13.1 REPLACING THE ULTRA-FINE FILTER
Part #362522
Included with the Replacement Kit:
• Ultra-Fine Filters (package of 6)
Electrical Shock Hazard: Disconnect the electrical supply before making repairs to the unit.
WARNING:
STEP 1
Remove the pollen filter by gently pulling from
around the edges.
STEP 2
Remove the plastic filter retaining ring, (if being used), by
pulling the tabs on the side and top of the ring out of the slots on
the unit. Remove the ultra-fine filter and discard. Replace with a
new filter. Check that the filter lays flat. Install the filter retaining
ring by bending the ring in the middle, (using the middle bar),
and inserting the ring tabs into the slots of the unit. Check that
the two side tabs are engaged into the unit slots.
STEP 3
Orient the pollen filter with the notched edge facing down and
gently tuck the edges in under the lip of the unit.
NOTE:
Ultra-fine filters should be replaced once per month or more frequently if necessary.
NOTE:
If using the optional ultra-fine filter, the filter retaining ring or filter securing kit must be used
at all times.
REMstar Service Manual
(REMstar)
Filter
Repair and Replacement
Top Tab
Middle Bar
Side Tab
6-47