Respironics REMstar Choice Service Manual page 99

Nasal cpap system
Table of Contents

Advertisement

STEP 3
Remove the time meter wires from the
cable clip on the rear panel of the unit.
Remove the wires from the wire tape
attached to the blower or the second
cable clip located on the rear panel,
whichever is applicable.
STEP 4
Spray isopropyl alcohol around the
bottom of the time meter, where it fits
through the opening in the unit base.
STEP 5
Using needle nose pliers, pull the glue bead from both sides of the time meter.
STEP 6
Lay the unit on the rear panel. Push the time meter through the bottom of the base, and remove
from the unit.
STEP 7
Remove any remaining hot melt adhesive from the base of the unit.
STEP 8
Place the new time meter into the opening and snap into place.
REMstar Service Manual
Repair and Replacement
Cable Clip
PUSH
Time Meter
Adhesive
Circuit
Board
6-29