Using a Direct Insertion Probe (DIP)
Analyzing Cholesterol with the Ion Trap Series

Step 11: Analyze the Data File

18
Ion Trap and DSQ Series Direct Probe System User's Guide
1. Wait until the DPC display panel indicates that the probe has cooled down below 100 °C.
2. Select the Insert/Remove Probe button again from the Status tab.
a. Select the Remove Probe tab and carefully follow steps 1–6 shown on the
Insert/Remove Probe screen.
Figure 16. Remove Probe Tab: Ion Trap Series
b. Click the Close button to return to the Sequence Setup window.
c. Begin the next step.
Use these steps to process the data file you just acquired. If you want to analyze another
cholesterol sample, repeat the steps listed
making the appropriate addition to the sequence table.
1. At the Sequence Setup window, click the Roadmap icon to return to the Xcalibur Home
Page.
Select Qual Browser to open the Qual Browser window. Refer to the Qual Browser
manual to learn how to use Qual Browser techniques.
2. Select File | Open and select the CholDIP.raw data file you just ran.
under"Step 6: Prepare the Sample"
on
page
12,