5
Calibration / Deskew Procedure

Procedure

1
To perform a DC calibration / calibrated skew, make sure the 50 ohm terminator (included
with N5443A) is connected to the fixture as shown below.
2
Hold the N5443A upright and connect the fixture to the Cal Out on the oscilloscope. Turn
the nut on the Cal Out counter-clockwise to tighten.
3
While holding the fixture upright with one hand, use an 8 in. lbs. torque wrench to fully
tighten the connector as shown on the next page.
76
Allow the probe to warm up for 20 minutes before performing the
N OT E
calibration procedure.
InfiniiMax III Series Probes User's Guide