Performing an exposure using a CR cassette

Topics:
Step 1: retrieve the patient info
Step 2: select the exposure
Step 3: prepare the exposure
Step 4: check the exposure settings
Step 5: execute the exposure
Step 6: repeat steps 2 to 5 for the next subexposures
Step 7: digitize the image
Step 8: perform a quality control
Valory (system with floor mounted column) | Basic Workflow | 91
Note: Using an ID Tablet to identify cassettes before the exposure
will break the communication of X-ray parameters between the NX
workstation and the X-ray generator console. It is advised to
identify cassettes after the exposure, as described in this workflow.