Step 1: retrieve the patient info

At the MUSICA Acquisition Workstation:
1. When a new patient comes in, define the patient info for the exam.
2. Start the exam.
If the workstation is connected to a second monitor that is positioned outside
the operator room, make sure that the patient data is not exposed to
unauthorized persons.

Step 2: select the exposure

In the operator room:
At the NX workstation, select the thumbnail for the exposure in the Image
Overview pane of the Examination window.
The default X-Ray exposure parameters for the selected exposure are sent
to the modality and displayed on the software console.
The selected DR Detector is activated.
The DR Detector Switch shows which DR detector is active and shows its
status.
• Flashing: starting up
• Green (constant): ready for exposure
Valory (system with floor mounted column) | Basic Workflow | 85