60 | Valory (system with floor mounted column) | Introduction
Do not open the equipment for cleaning. No components inside the device
require cleaning by the user.
4. Wipe the exterior of the system with a dry low-linting cloth or a cloth
slightly moistened with water, to remove all cleaning agent residue.
Let all surfaces dry.
5. Perform a visual inspection of the exterior of the system.
The room must be sufficiently illuminated to perform the visual
inspection.
Check for deterioration such as corrosion, cracks or serious discolorations,
for damage to the cables or cable jackets, for loos screws.
Make sure that the accessories for immobilizing the patient haven't come
loose.
If there's still dirt or cleaning agent residue, repeat the cleaning.
6. Disinfect all parts of the system which are contaminated or frequently
touched by the user or the patient.
1.
2.
3.
4.
Figure 39: How frequently are parts of the system touched?
0426A EN 20220804 1858
unsuitable cleaning agents or methods can damage the
property when surface becomes dull and brittle (e.g. by
using alcohol-containing agents).
Very often
Regularly
Rarely
Almost never