188 | Valory (system with floor mounted column) | Product Information
MUSICA Ac-
quisition
workstation
(typical,
without UPS
and extra
monitors)
High voltage output values
High voltage
power out-
put (at 0.1s)
kV-Range
mAs-Range
mA-Range
ms-Range
Mechanical data
Dimensions
Weight
Duty cycle
(*) Split phase AC line with two hot phases, e.g. 240 VAC available as two 120
VAC hot phases
0426A EN 20220804 1858
45 W
500mA: 80kV
400mA: 100kV
320mA: 125kV
266mA: 150kV
40-150 kV
0.5-600 mAs
10-500 mA
1-6300 ms
550 mm x 630 mm x 970 mm (WxDxH)
129 kg
The Generator duty cycle is continuous, but limits should be
set during installation depending on the capacity of the X-ray
tube.
625mA:
800mA:
80kVp
80kVp
500mA:
650mA:
100kVp
100kVp
400mA:
520mA:
125kVp
125kVp
330mA:
430mA:
150kVp
150kVp
10-650 mA
10-800 mA
800mA:
80kVp
800mA:
100kVp
640mA:
125kVp
530mA:
150kVp