108 | Valory (system with floor mounted column) | Software Console

Radiation status

Table 17: Radiation status
Press the exposure button halfway ("Prep" position) to prepare the X-ray tube
for exposure. The indicator will light up when the X-ray tube is prepared and
there are no interlock failures or system faults.
After pressing this push-button, the following functions are activated:
• Anode rotation.
• Filament current switches from stand-by to the selected mA.
0426A EN 20220804 1858
The X-ray tube is prepared.
After pressing the exposure button completely, the X-ray exposure is
made. The indicator on the console will light up.
The examination room door is open.