66 | Valory (system with floor mounted column) | Introduction
1. On the software console, select the manual modality positon.
No image will be acquired on the NX workstation.
2. Select the three point radiographic working mode.
3. Set the radiographic parameters to 125 mA (current) and 100 ms
(exposure time).
4. Select the large focal spot.
5. Take a sequence of exposures with the following kV values. Take one
exposure per 30 seconds.
Table 4: Sequence of exposures
Time (minutes)
0.0
0.5
1.0
1.5
2.0
2.5
3.0
3.5
0426A EN 20220804 1858
mAs
mA,ms
kV Time (minutes)
50 4.0
50 4.5
60 5.0
60 5.5
70 6.0
70 6.5
80 7.0
80 7.5
kV
Time (minutes)
90
8.0
90
8.5
100 9.0
100 9.5
110 10.0
110 10.5
120
120
kV
130
130
140
140
150
150