Valory (system with floor mounted column) | Basic Workflow | 83
Figure 40: Navigation button for modality controls
4. Click the button to start the automated workflow for warming-up of the X-
ray tube.
Figure 41: Button to start the automated workflow for warming-up of the
X-ray tube
A table is displayed with a list of exposures.
1
xx
xx
xx
xx
xx
xx
xx
xx
2
Table with list of exposures
1.
Button to cancel the warming-up procedure
2.
Figure 42: List of exposures for warming-up of the X-ray tube
5. Make sure that the collimator blades are fully closed and that no patient is
present in the room.
To avoid radiation on a DR detector, remove the detector, turn the tube
away from the detector, or cover the detector with a lead apron.
6. Perform the exposures and wait for the timer icon to finish between
exposures.
The exposure parameters are set automatically.
xx
xx
xx
xx
0426A EN 20220804 1858