194 | Valory (system with floor mounted column) | Product Information
Exposure time [ms]
Step
0
1
2
3
4
5
6
7
8
9
Maximum tube current [mA] at 100 kVp and 0.1 s
E7884X
E7252X
E7254FX
• LSS: Low Speed Start option
• HSS: High Speed Start option
All Values are valid for 3-phase generator power line and large focal spot.
Values for other exposure conditions can be determined using the technical
data of the generator and the data sheets of the X-ray tubes.
0426A EN 20220804 1858
ms
Step
ms
1
10
13
2
11
16
3
12
20
4
13
25
5
14
32
6
15
40
7
16
50
8
17
63
10
18
80
11
19
100
Note: Not all exposure parameters may be available, depending
on the configuration of X-ray generator, X-ray tube and DR
Detector.
HFe 401
HFe 501
(40 kW)
(50 kW)
LSS: 400 mA
LSS: 500 mA
LSS: 400 mA
LSS: 450 mA
HSS: 400 mA
HSS: 500 mA
LSS: 400 mA
LSS: 500 mA
HSS: 400 mA
HSS: 500 mA
Step
ms
20
130
21
160
22
200
23
250
24
320
25
400
26
500
27
630
28
800
29
1000
HFe 601
(65 kW)
-
HSS: 650 mA
HSS: 650 mA
Step
ms
30
1250
31
1600
32
2000
33
2500
34
3200
35
4000
36
5000
37
6300
HFe 801
(80 kW)
-
-
HSS: 800 mA