AGFA Valory User Manual page 75

System with floor mounted column
Table of Contents

Advertisement

Valory (system with floor mounted column) | Safety Directions | 75
WARNING:
Make sure that no clothes of the patient or the operator get
caught by moving parts of the system.
WARNING:
Portable and mobile HF communication devices may affect
medical electrical equipment.
CAUTION:
Excessive ambient temperature may impact performance of DR
Detectors and cause permanent damage to the equipment. Refer
to the related user manual for environmental conditions for the
DR detector. If ambient temperature and humidity is outside the
specified range, do not operate the system or use air
conditioning. Frost due to low temperatures can damage
internal circuits. Warranty will be void if it is obvious that
operating conditions are not met.
CAUTION:
Sudden heating of the room in cold areas will cause
condensation to form on the equipment. In this case, wait until
the condensation evaporates before use. If the equipment is
used while condensation is formed on it, problems may occur.
Condensation inside the equipment can cause rust and
corrosion. When an air-conditioner is used, be sure to raise/
lower the temperature gradually so that a difference in
temperature in the room and in the equipment does not occur,
to prevent condensation.
CAUTION:
To avoid images being lost due to a power failure, use an
uninterruptable power supply (UPS) or an institutional standby
generator.
WARNING:
System unavailability due to hardware or software failure. If the
product is used in critical clinical workflows, a backup system
has to be foreseen.
CAUTION:
Due to the length of high voltage cable, the X-ray tube head unit
should not be rotated over more than ±180°.
CAUTION:
Strictly observe all warnings, cautions, notes and safety
markings within this document and on the product.
0426A EN 20220804 1858