Expected lifetime for the X-ray unit
Periodic maintenance
An Agfa trained and authorized service engineer
shall perform a technical maintenance according to
the maintenace schedules in the service documen-
tation, to maintain fault-free operation and ensure
safety for patient and operator.
Maintenance by the user
Check the room for safety and remove objects that
are within the movement area of the system
Check constant smooth movements
Check ease of movements
Check secure release and locking of brakes
Check functioning of operating controls
Check markers and warning signs
Check the equipment for visual damage, deforma-
tion or dents
Check all electric cables and connections for mois-
ture and make sure they're dry
Check all electric cables and connections for dam-
age or broken cables.
Warm-up of X-ray tube
Conditioning of the X-ray tube
Conditioning of the X-ray tube
Conditioning procedure for the X-ray tube
If the X-ray tube has not been used for more than a week or if exposure
techniques are to be used with energies above 120 kV, it is recommended to
perform the conditioning procedure for the X-ray tube.
A sequence of gradually increasing loads on the X-ray tube will cause a
redistribution of the electrical charges inside the tube, which in turn will
result in a stable output of the tube.
The procedure takes approximately 30 minutes.
Valory (system with floor mounted column) | Introduction | 65
10 years
Every 12 months
Daily
Daily
Daily
Daily
Daily
Daily
Daily
Daily
Daily
Daily
After the X-ray tube
has not been used for
more than a week
Before making expo-
sures using voltages of
120 kV or higher
0426A EN 20220804 1858