ECOTEST MKS-UM Operating Manual page 78

Multipurpose dosimeter-radiometer
Table of Contents

Advertisement

2.3.6.10 Working with measurement results and DE accumulation
history
2.3.6.10.1 General information
Measurement results recorded in the nonvolatile memory of the control unit
can be viewed on its LCD, and communicated for further processing to the personal
computer. DE accumulation history can be also viewed on the LCD of the control
unit. You can work with DE accumulation history only using the PC.
The PC should be equipped with the USB/IrDA adapter manufactured by PE
"SPPE "Sparing-Vist Center" and it should have the installed custom-made software.
2.3.6.10.2 Operating mode of the control unit "Measurement results
viewing saved in the nonvolatile memory"
2.3.6.10.2.1 The mode of measurement results viewing saved in the
nonvolatile memory is available only when the nonvolatile memory of the control
unit contains at least one saved measurement result. This mode can be accessed from
any other mode by shortly pressing MODE. This mode follows the mode of current
time display.
2.3.6.10.2.2 In this mode the LCD of the dosimeter displays the following (Figure
2.37):
symbol (1) – indication of the mode of measurement results viewing
-
saved in the nonvolatile memory;
- measurement result (2);
- indicator of measurement result location in the nonvolatile memory (3);
- measurement object number (4);
-
symbol - indication of the availability of geographical coordinates of the
measurement object in the memory (5);
- battery status symbol (6).
Important!
part of the oldest information was lost (2.3.4 of this OM).
(Mode of measurement results viewing saved in the nonvolatile memory)
symbol (1) is blinking, if during operation of the dosimeter a
Figure 2.37 – LCD of the dosimeter
76

Advertisement

Table of Contents
loading

Table of Contents