Quantum Data 881 User Manual page 113

Video test generator
Hide thumbs Also See for 881:
Table of Contents

Advertisement

881/882 Video Test Generator User Guide (Rev A.25)
2. Open a second Web browser. You now have two instances of the Generator FTP
Browser running: one for the source generator and one for the target generator.
3. In the Generator FTP Browser window for the target generator, delete all of the files in
the target generator's flash memory as follows:
a. In the Instrument Files area, click the down arrow by the Look in box and select
tffs0. This is the target generator's flash memory.
b. Select all of the System and Library folders in the lower pane.
c. Click the Delete button. A confirmation dialog box appears.
d. Click OK to delete the files.
4. Copy all of the files in the source generator's flash memory to the target generator's
flash memory as follows:
a. In the Instrument Files area of source generator's Generator FTP Browser
window, click the down arrow by the Look in box and select tffs0. This is the source
generator's flash memory.
b. Select the System and Library folders in the lower pane.
c. Hold down the SHIFT key and drag the selected folders to the Instrument Files
area of the target generator's Generator FTP Browser window. A confirmation
dialog box appears.
d. Click OK to copy the files.
111

Advertisement

Table of Contents
loading

This manual is also suitable for:

882

Table of Contents