<NR1>

STATus:OPERation:CONDition?

Queries the condition register for the Operation Status group. This is a read-only
register, which holds the live (unlatched) operational status of the instrument.
Reading the Operation Status Condition register does not clear it.
Group
STATus
Syntax
STATus:OPERation:CONDition?
Arguments
None
Default Value
None
Returns
<bit value>
Example
STATus:OPERation:CONDition?
The value returned is the binary-weighted sum of all enabled bits in the
register. For example, with bit 3 (value 8) and bit 5 (value 32) set and en-
abled, the query returns +40.
The condition register bits reflect the current condition. If a condition goes
away, the corresponding bit is cleared.
*RST clears this register, other than those bits where the condition still exists
after *RST.