Source:]Function:mode <Cpd - ITech IT8000 Series Programming Manual

Regenerative dc electronic load
Hide thumbs Also See for IT8000 Series:
Table of Contents

Advertisement

Example
LOOP:SPEed HIGH
Query Syntax
[SOURce:]LOOP:SPEed?
Returns
HIGH|LOW
12.3 [SOURce:]FUNCtion:MODE <CPD>
This command is used to set the functional mode of the load, i.e. transient
mode. This determines what happens to the input current when the transient
system is initiated and triggered.
Group
SOURce
Syntax
[SOURce:]FUNCtion:MODE <CPD>
Arguments
<CPD>
FIXed|ARB|BATTery
Default Value
FIXed
FIXed: The default value indicates that the instrument is operating in fixed
mode. Keeps the input current at its immediate value.
ARB: Indicates that the instrument is operating in ARB mode. When a trigger
occurs, ARB will cause the input to follow the list value.
BATTery: Indicates that the instrument is operating in battery test mode.
When a trigger occurs, it is input based on the edited battery test file.
Copyright © Itech Electronic Co., Ltd.
SOURce Subsystem
124