contact sales for advice)
required for the DCS to communicate with the Ai1 and for the Ai1 to decode results correctly.
STOP
GET ALL
HELP
SET PASSKEY(1)
SET MODE(SMART SENSOR TERMINAL)
SET ENABLE DECIMALFORMAT(YES)
SET ENABLE TEXT(NO)
SET PING NUMBER(50)
SET POLLED PINGRATE(10.0)
SET ENABLE POLL MODE(NO)
SET INTERVAL(10)
SAVE
RESET
The DCS does not have a function to reset to factory defaults so it is best to check each configuration parameter
listed above for correct values. For a listing of all factory defaults, a file is available for download:
www.measci.com/dcs_default.txt
The Ai1 controls power to the sensor so that it is only turned on when required, and off in-between readings. The
readings returned by the DCS which may be displayed in Hydrosphere, Eagle.io or other platforms list in Table 9.
Table 9 - DCS Parameter Names
Parameter Name
DCS_Model
DCS_SN
Abs_Speed
Direction
Nth_Vector
Eth_Vector
Heading
Tilt_X
Tilt_Y
SP_Std
Sig_Strength
Ping_Count
Abs_Tilt
Max_Tilt
Std_Tilt
DCS_Temp_ZPulseOnly Temperature (only fitted on Z pulse models)
#Do I have decimal mode on or not? It displays zero as 0.0000000E+00. I think I should have it on.
14
To set up for communication with a PC, connect Green to DB9 pin 2, White to DB9 pin 3 and Black to Ai1 GND. DCS remains powered by Ai1. The terminal must be set up with CR+LF line feeds.
Page | 50
14
. To set up a DCS, the following settings are recommended. Items highlighted bold are
If the sensor is currently running measurements. This may need to be
repeated
Show all the available settings
Show all the available commands. To see all modes with an
enumerated parameter, view the options under "Argument List for
enumerated properties" at the end of the help message
Passkey level 1 gives permission to write most parameters, although
some may require Passkey level 1000. If a set command results in the
message "ERROR PROTECTED PROPERTY", a higher passkey is required.
Changes between modes such as AiCaP and the serial terminal mode.
Display values in decimal format, i.e. 0.000 instead of 0.000E+00
Set data reports to be numbers only, no text headers
The number of pings for a reading, options 50, 60, 75, 100, ..., 600
The number of pings per second, options 0.1, 0.5, 1.0, 2.0, 5.0, 10.0,
15.0, 20.0, 25.0. Note it must include the decimal. The result of Ping
Number/Ping Rate should be less than 10 so that one reading is
returned each 10s
Does not require polling, transmits based on an interval
The number of seconds between data output reports
Save the settings to non-volatile memory
Run the sensor with the new settings
Description
DCS Model, i.e., 4830
DCS Serial Number
Average Absolute Current Speed (vector averaged)
Current Direction of Abs_Speed vector
North component of current direction
East component of current direction
Last measured compass heading
Tilt in X direction of DCS
Tilt in Y direction of DCS
Single Ping standard deviation of all pings in measurement
Signal Strength
Number of pings used in measurement
Absolute tilt (relative to horizontal plane) of DCS
Maximum tilt (relative to horizontal plane) of DCS
Standard deviation of tilt measurements
Default Units
cm/s
Degrees (°)
Degrees (°)
Degrees (°)
Degrees (°)
Degrees (°)
Degrees (°)
cm/s
dB
pings
Degrees (°)
Degrees (°)
Degrees (°)
°C