Inspector 5000™ Auto Optic
Operator's Guide
Press Up button to display the reflectometer mode and to
verify that the unit is calibrated within tolerance placing the
Auto Optic on the light or dark area of the plaque.
Reflectance values should be within +/- 2% (of the value
written the calibration plaque – "High Reflectance Value
Target Values: Red (660nm)") for white and +/- 2% (of the
value written the calibration plaque – "Low Reflectance
Value Target Values: Red (660nm)") for black.
RJS, Minneapolis, MN
003-4500
18