SECTION 9: SPECIFICATIONS

9.1

GENERAL

9.2

ELECTRICAL

9.2.1 Protection Class
9.2.2 Degree of Protection
9.2.3 Mode of Operation
9.2.4 Battery
9.2.5 Fuses
9.2.6 AC Power
9.1 General
9.2 Electrical
9.3 Physical Characteristics
9.4 Environmental
9.5 Alarms
9.6 Factory Default Settings
9.7 Performance
Designed to meet safety requirements of:
UL 2601-1 CSA-C22.2 No. 601-1-M90, IEC 601-1 (Class I, type BF), ISO 9919,
EMC per EN 60601-1-2.
Class I: per I.E.C. 601-1, clause 2.2.4
Type BF: per I.E.C. 601-1, clause 2.2.26
Continuous
Type:
Operating time:
Recharge period:
Selectable by switch
Rechargeable sealed lead-acid, internal
6 hours minimum on new, fully charged
battery with no backlight or alarms
14 hours for full charge (in standby)
18 hours for full charge (in use)
2 each 5 x 20 mm
0.5 Amp 250 volts
100-120 VAC 50/60 Hz or
200-240 VAC 50/60 Hz