Section 6: Disassembly Guide
6.5

MONITOR REASSEMBLY

6-4
Figure 6-3: Separating Case Halves
1.
Connect the Power Supply to J6 on the UIF PCB.
2.
Place the top case over the bottom case, being careful to align the lens,
Power Entry Module, and the fan with the slots in the top case.
Caution: When reassembling the NPB-195, tighten the screws that hold the
cases together to a maximum of 10 inch-pounds. Over-tightening could
strip out the screw holes in the top case, rendering it unusable.
3.
Install the four corner screws.