Section 3: Performance Verification
3.4.2.4
Patient Isolation Risk Current - (Mains Voltage on the Applied Part)
3-12
This test is in compliance with AAMI Standard ES1 (patient isolation risk
current [sink current]), and IEC 601-1 (patient leakage current). Patient Leakage
Current is the measured value in a patient connection if mains voltage is
connected to that patient connection. The applied voltage for AAMI/ANSI is
120 VAC 60 Hz, and for IEC 601-1 the applied voltage is 264 VAC 50 to 60 Hz.
Warning: AC mains voltage will be present on the patient applied part
terminals during this test. Exercise caution to avoid electrical shock hazard.
1.
Configure the electrical safety analyzer as follows:
Function:
Range:
2.
Connect the monitor AC plug to the electrical safety analyzer as
recommended by the operating instructions for patient sink (leakage)
current.
3.
Connect the patient leakage input lead to the electrical safety analyzer to all
connectors in the patient cable at the patient end of the cable.
4.
The equipotential terminal is not connected to ground.
5.
All functional earth terminals are not connected to ground.
6.
The analyzer leakage current must not exceed the values shown in
Table 3-5.
Table 3-5: Patient Leakage Current Test Configurations -
Mains Voltage on the Applied Part
AC LINE
NEUTRAL
POLARITY
LINE
Normal
Closed
Reverse
Closed
Patient Leakage (Mains On Applied Part)
µA
POWER
LINE
GROUND
CABLE
Closed
Closed
IEC 601-1
AAMI/ANSI
ES1
STANDARD
5 mA
50 µA
5 mA
50 µA