Service Functions; Table 4-1: Factory Default Settings - Mallinckrodt NELLCOR NPB-195 Service Manual

Table of Contents

Advertisement

Section 4: Power-On Settings and Service Functions
4.3

SERVICE FUNCTIONS

4.3.1 Introduction
4.3.2 Accessing the Service Functions
4-2

Table 4-1: Factory Default Settings

Parameter
SpO
High
2
SpO
Low
2
Pulse Rate High
Pulse Rate Low
Alarm Volume
Alarm Silence Duration
Alarm Silence Restriction
Pulse Beep Volume
Baud Rate
Display
Trend
Contrast
Language
Service functions can be used to select institutional defaults, and to access
information about the patient or instrument. Only a Mallinckrodt Customer
Service Engineer should access some of the items available through the service
functions. These functions will be described in the text to follow.
All service functions are accessible only if the sensor cable is disconnected from
the instrument. Simultaneously press the 4th softkey from the left and the
contrast button for more than 3 seconds. The menu bar will change to the
headings listed in Figure 4-1.
Note: If a "Sensor Disconnected" prompt appears on the screen, press the
Alarm Silence button and repeat the above procedure.
Note: If the above steps are performed with a sensor cable connected, only the
Param and Exit softkeys appear on the screen.
Figure 4-1: Service Function Softkeys
Default Value
100%
85%
170 bpm
40 bpm
Level 5
60 seconds
Sound Reminder
Level 4
9600
Pleth
Saturation
Mid-range
English
%SP02
- - -
- - -
BPM
PARAM
PRINT
NEXT
EXIT
NPB-195