Specifications - Mallinckrodt INOblender 10026 Operation And Maintenance Manual

Inoblender series
Table of Contents

Advertisement

Specifications

INOblender
Ü NO Set Range
Setting accuracy
Gas flow range
O
inlet connection
2
NO/N
inlet pressure
2
NO inlet
NO/O
outlet
2
Max. Dimensions
Max. weight
Operating and Storage Conditions
Operating Temperature
Operating Humidity
Operating Barometric Pressure
Storage Temperature
Storage Humidity
Storage Barometric Pressure
INOMAX Pressure Regulator
Cylinder Valve Connector
Inlet Pressure
NO Outlet Pressure
Manual Resuscitators
The INOblender is compatible with most types of manual resuscitation bags when used within the specifications of this
device. The INOblender has been validated against the following manual resuscitation bags.*
* Airlife Adult Manual Resuscitator with reservoir tubing
* Allegiance ½ L Pediatric Manual Resuscitator with reservoir tubing
5 to 80 ppm (800 ppm cylinder)
2.5 to 40 ppm (400 ppm cylinder)
+/- 20 % indicated or 2 ppm, whichever is greater
5 to 14 L/min of O
2
DISS (male), 3.4 bar (50 psig, nominal)
1.7 to 2.4 bar (25 to 35 psig)
Quick disconnect with shut-off (Female)
Barb fitting for oxygen supply tubing
200 mm (W) X 120 mm (H) X 110 mm (D) DISS (clamp included)
200 mm (W) X 120 mm (H) X 131 mm (D) NIST (clamp included)
1.7 kg (clamp included)
10 to 40°C
15% to 95% non-condensing
57 to 110 kPa
-40 to + 70°C
15% to 98% non condensing
57 to 110 kPa
CGA 626 or ISO 5145 valve
14 to 155 bar (200 to 2,248 psig)
1.7 to 2.4 bar (25 to 35 psig)
21
Part No. 20732 Rev-02
2014-02