Section 4: Power-On Settings and Service Functions
4-6
NPB-195 Version 1.0.0.000
Op Time
Error
10713:21:03
52
00634:26:01
37
Output Complete
INSTAT
A Delete softkey, described in the Operator's manual, allows the user to clear the
most recent trend data. Cleared trends can still be retrieved from the instrument
through an Instat printout. Instat can be accessed by pressing Next after the Print
softkey has been selected.
The oldest cleared trend will become Trend 01 on the Instat printout. If a Trend
01 already existed in memory from an earlier Clear, the next clear will become
Trend 02. Every time a Clear is performed from the User Softkeys, the number
of existing trends will increase by 1.
Figure 4-7 illustrates an Instat printout. The first line is for instrument type,
software revision level, type of printout, and alarm parameter settings. The
second line lists the column headings. A line of data is recorded for every 5
seconds of instrument operation. Up to 24 hours of instrument operation data
can be recorded.
The final line on the printout is Output Complete. This indicates that data has
been successfully transmitted with no corruption. If no Output Complete line is
printed, the data should be considered invalid.
NPB-195
Version 1.0.0.000
TIME
Trend 01
%SpO2
01-Jul-97 14:00:00
- - -
01-Jul-97 14:00:05
- - -
01-Jul-97 14:00:10
100
01-Jul-97 14:00:15
100
NPB-195
Version 1.0.0.000
TIME
Trend 01
%SpO2
01-Jul-97 14:24:24
79*
01-Jul-97 14:24:29
79*
01-Jul-97 14:24:29
0*
NPB-195
Version 1.0.0.000
TIME
Trend 01
%SpO2
11-Jul-97 7:13:02
99
11-Jul-97 7:13:07
99
11-Jul-97 7:13:12
99
11-Jul-97 7:13:17
99
11-Jul-97 7:13:22
99
11-Jul-97 7:13:27
99
11-Jul-97 7:13:32
99
Output Complete
Error Log
Task
Addr
12
48F9
4
31A2
Figure 4-6: Errlog Printout
Instrument
SpO2 Limit:
PR (bpm)
PA
SpO2 Status
- - -
- - -
SD
- - -
- - -
120
220
120
220
Instrument
SpO2 Limit:
PR (bpm)
PA
SpO2 Status
58*
220
PS
57*
220
PS
0*
- - -
PS LP SL PL
Instrument
SpO2 Limit:
PR (bpm)
PA
SpO2 Status
132*
220
132*
220
132*
220
132*
220
132*
220
132*
220
132*
220
Figure 4-7: Instat Printout
Time:
14600:00:07
Count
100
3
30-100%
PR Limit: 100-180 bpm
UIF Status Aud
BU LB
PS
BU LB
BU LB
BU LB
80-100%
PR Limit: 60-180 bpm
UIF Status Aud
SL PL
BU LB
SL PL
BU LB AS
BU LB AS
80-100%
PR Limit: 60-180 bpm
UIF Status Aud
PH
BU
PH
BU
PH
BU
PH
BU
PH
BU
PH
BU
PH
BU
AO
L
AO
M
M
H
M
M
M
M
M
M
M