3.3.3.1.2
Alarm Volume Control
3.3.3.1.3
Pulse Tone Volume Control
3.3.3.1.4
Dynamic Operating Range
After completing the procedure in paragraph 3.3.3.1.1 above:
1.
Press and hold the Alarm Silence button and verify the following:
a.
"OFF" is displayed for approximately 2 seconds.
b.
After 2 seconds, a steady tone is heard at the default alarm volume
setting, the %SpO
indicates the default setting of 5.
2.
While still pressing the Alarm Silence button, press the Down Arrow button
until an alarm volume setting of 1 is displayed. Verify that the volume of
the alarm has decreased but is still audible.
3.
Continue pressing the Alarm Silence button and press the Up Arrow button
to increase the alarm volume setting to a maximum value of 10. Verify that
the volume increases. Press the Down Arrow button until a comfortable
audio level is attained.
4.
Release the Alarm Silence button. The tone will stop.
1.
Press the Up Arrow button and verify that the beeping pulse tone sound
level increases.
2.
Press the Down Arrow button and verify that the beeping pulse tone
decreases until it is no longer audible. Press the Up Arrow button to return
the beep volume to a comfortable level.
The following test sequence verifies proper monitor operation over a range of
input signals.
1.
Connect the SRC-2 to the NPB-195 and turn the NPB-195 on.
2.
Place the SRC-2 in the RCAL 63/LOCAL mode.
3.
Set the SRC-2 as indicated in Table 3-1. Verify that the NPB-195 readings
are within the indicated tolerances. Allow the monitor several seconds to
stabilize the readings.
Note:
A asterisk "*" indicates values that produce an alarm. Press the Alarm
Silence button to silence the alarm.

Table 3-1: Dynamic Operating Range

SRC-2 Settings
RATE
LIGHT
38
HIGH2
112
HIGH1
201
LOW
201
LOW
Section 3: Performance Verification
display indicates "VOL," and the Pulse display
2
MODULATION
SpO
LOW
79 - 83* 35 - 41*
HIGH
79 - 83* 109 - 115
LOW
79 - 83* 198 - 204*
HIGH
79 - 83* 198 - 204*
NPB-195 Indications
Pulse Rate