Troubleshooting

Problem
Continuous
beeping sound
Low EtCO
alarm
2
even though the
patient's EtCO
is
2
suspected to be
normal.
WARM UP with
alarms
28
Possible Cause
The alarm beeps continuously.
The monitor is not functioning. This
indicates that a problem has
occurred, possibly due to
interference or loss of power.
All alarms for low EtCO
operator to check the patient's status.
It is also possible to get a low reading
if an air leakage has occurred in the
sample line, Nafion tubing, moisture
trap, or internally.
All abnormal readings have to be
checked with respect to the patient's
condition. If the readings are out of
range, one may also suspect an
equipment fault.
Turn off the monitor and then turn
on again.
Recharge the monitor with the
power supply.
If the problem persists, contact
Nonin Technical Service.
require the
Check patient status.
2
Check the moisture trap and filter.
Replace the moisture trap and
filter if necessary.
Check sample line connector and
visually inspect the sample line for
signs of damage.
Check Nafion tubing connection.
If the problem persists, contact
Nonin Technical Service.
Verify the filter is in place. Replace
as needed.
Perform calibration and gas
verification to assure performance
of the device.