Indications for Use

RespSense is a lightweight, portable, battery-operated monitor that measures and displays
carbon dioxide in expired air (EtCO
It is intended for use in environments where patients require continuous, non-invasive monitoring
of these parameters by a healthcare professional (e.g., hospitals, medical facilities, post-
operative care, within facility patient transport, and home use.

Warnings

Do not use RespSense in an MR environment or in the presence of flammable anesthetics or gases.
Do not use RespSense during defibrillation.
RespSense is intended only as an adjunct in patient assessment. It must be used in conjunction with other
methods of assessing clinical signs and symptoms.
RespSense is not intended to be used as a primary diagnostic apnea monitor.
Verify all alarm settings and limits during system start up to ensure that they are set as intended.
A hazard can exist if different presets are used on multiple RespSense monitors in one care area.
To prevent improper performance and/or injury to the patient, verify compatibility of the monitor and accessories
before use.
No modifications to this device are allowed as it may affect device performance.
Never allow liquids to enter into or to be spilled onto the monitor. If liquid has penetrated into the monitor, it must
be checked by Nonin Technical Service.
Accessories marked "single-use" must be used on one patient only and be disposed of after usage.
Use only power supplies that either are supplied with RespSense or specified by Nonin (see "Accessories").
If RespSense fails to respond as described, discontinue use and contact Nonin Technical Service.
Use only Nonin-recommended accessories and replacement parts.
RespSense displays a BATT LOW message when it has approximately 20 minutes of use remaining before it
shuts itself off.
EtCO
value will be diluted when used in combination with supplemental oxygen. To get a true EtCO
2
is recommended that the supplemental oxygen is disconnected for a few seconds.
This device should not be used adjacent to or stacked with other equipment. If adjacent or stacked use is
necessary, the device should be observed carefully to verify normal operation.
Ensure that all alarm volumes are audible in all situations. Do not cover or obstruct any speaker openings.
Prior to connecting RespSense to the power supply and the power outlet, be sure to verify the voltage and
frequency rating on the power supply are the same as the outlet. If this is not the case, do not connect the
monitor and power supply to the outlet.
When turning on the monitor, verify that a beep is heard each time a button is pressed. If a beep is not heard, do
not use the device. The speaker may not be functioning properly.
), and respiration rate of adult, pediatric, and infant patients.
2
reading, it
2
1