Troubleshooting - Nonin 9847V Operator's Manual

Veterinary handheld pulse; oximeter and carbon dioxide (co2) detector
Table of Contents

Advertisement

Troubleshooting

Troubleshooting
Problem
The device won't turn on.
The Low Battery indicator is
steadily lit or flashing.
A dash appears in the
leftmost position of the SpO
display.
The middle digits display
dashes in both the SpO
Pulse Rate displays.
The displayed pulse rate
does not correlate to the
pulse rate displayed on the
ECG monitor when used
together.
An inconsistent pulse rate or
a yellow Pulse Quality
indicator during use with
electrosurgical unit (ESU).
The Pulse Quality indicator
blinks yellow with each
pulse.
Possible Cause
Batteries are depleted.
Batteries are installed incorrectly.
Battery level is low or critically low.
Incorrect battery installation.
A SpO
sensor fault exists
2
(disconnect, failure, misalignment, or
2
incompatibility with the monitor).
A non-compatible SpO
being used.
No SpO
signal is detected.
2
and
2
A sensor failure.
Excessive motion at sensor site may
be prohibiting device from detecting a
consistent pulse signal.
Animal may have an arrhythmia
resulting in some heart beats that do
not detect a pulse quality signal at the
sensor site.
Non-compatible sensor is being
used.
ECG monitor may not be functioning
properly.
ESU may be interfering with pulse
oximeter performance.
Quality of pulse signal
at sensor site is inadequate.
Replace all six batteries.
Verify battery orientation, as
illustrated inside battery
compartment or on device label.
Replace all six batteries.
Verify correct battery orientation.
Verify Nonin-branded sensor is
correctly connected to the device
and the animal; replace sensor if
condition persists.
sensor is
Replace sensor with a Nonin-
2
branded Purelight sensor.
Verify sensor connection.
Replace sensor with a Nonin-
branded Purelight sensor.
Eliminate or reduce the cause of the
motion or reposition sensor to a new
sensor site.
Assess the animal.
Replace sensor with a Nonin-
branded Purelight sensor.
Assess the animal.
Assess the animal. Move device,
cables, and sensors as far away
from the ESU as possible.
Assess the animal. Reposition
sensor or select an alternate sensor
site.
Possible Solution
43

Advertisement

Table of Contents
loading

Table of Contents