Alarm Limits
All parameters have built in limits that cannot be exceeded.
Respiration Limits
EtCO
Limits
2
Changing Settings
All settings follow the same procedure to increase or decrease an alarm limit.
• The up arrow
on the right side of a displayed parameter bar is used to increase an alarm
limit.
• The down arrow
alarm limit.
Each time the arrow is pressed, it increases or decreases the alarm limit by a single digit until
the maximum or minimum is reached. The display scrolls through the values if the arrow is
steadily pressed.
The upper alarm limit is always located above the displayed value, and the lower limit is always
located below the displayed value.
Decreases upper limit
Decreases lower limit
CAUTION: Set or adjust only one parameter at a time.
!
NOTE: The monitor will always reset the alarm limits to the factory default settings once it is
turned off and turned on again.
20
– Upper limit: 99 RPM
– Lower limit: 0 RPM
– Upper limit: 9.9 kPa or 99 mmHg
– Lower limit: 0 kPa or mmHg
on the left side of a displayed parameter bar is used to decrease an
Increases upper limit
Increases lower limit